1.0 OBJECTIVETo describe a procedure for Calibration of Micropipette by using water.
2.0 SCOPE
This procedure is applicable to Calibration of Micropipette and is applicable in Microbiology
3.0 RESPONSIBILITY
3.1 Microbiologist /Officer is responsible for the execution.
3.2 HOD or designee is responsible for review, effective implementation and compliance to the calibration of micropipette for execution.
4.0 PROCEDURE

4.1 Operation
4.1.1 Set the required volume by rotating color knob clockwise and anti clockwise.
4.1.2 Fix the micro tip by applying light pressure.
4.1.3 Press the Plunger and insert the micro tip in the solution to be taken.
4.1.4 Release the plunger to get the required solution.
4.1.5 Transfer the solution by pressing the plunger little more than the normal. Press the dispenser knob to remove the used micro tip.
Note: Use this pipette (100 -1000µl) for volume not less than 100 µl and not more than 1000 µl Use this pipette (1 -10ml) for volume not less than 1ml and not more than 10ml.
4.2 Calibration
4.2.1 Set the volume at 200 µl, 500 µl and 1000µl for micropipette (T -1000µl).
4.2.2 Follow the operation procedure from point no 6.1.2 to 6.1.5.
4.2.3 Take the water at room temperature as per set volume and weigh it in a weighing bottle, which is previously dried and weighed.
4.2.4 Record the reading.
4.2.5 Repeat the analysis for 5 times and take the average.
4.2.6Calculate the volume by dividing the average weight of water with the actual volume from the table given as per the test temperature and record the values in Attachment #1
Temperature |
Volume of 1 gm of water |
Temperature |
Volume of 1 gm of water |
15 0 c |
1.0020 ml |
24 0 c |
1.0037 ml |
16 0 c |
1.0021 ml |
25 0 c |
1.0039 ml |
17 0 c |
1.0023 ml |
26 0 c |
1.0042 ml |
18 0 c |
1.0025 ml |
27 0 c |
1.0045 ml |
19 0 c |
1.0026 ml |
28 0 c |
1.0047 ml |
20 0 c |
1.0028 ml |
29 0 c |
1.0050 ml |
21 0 c |
1.0030 ml |
30 0 c |
1.0053 ml |
22 0 c |
1.0032 ml |
31 0 c |
1.0056 ml |
23 0 c |
1.00345 ml |
32 0 c |
1.0059 ml |
4.2.7 Acceptance criteria:
For micropipette 100µl –1000µl
Set volume in Permitted Volume
200-µl 196-204 µl
500-µl 495-505 µl
1000-µl 990-1010µl
Frequency – Quarterly
4.2.8 Precautions :
4.2.8.1 After fixing volume of micropipette don’t rotate the push button.
4.2.8.2 No foreign particle should be present in the tip as this can block the delivery of the liquid and this can result in incorrect volume dispenses.
0 Comments